Fluke 7009 /7108/ 7015 Resistor Baths

Features

Three size options for any quantity of resistors

  • Stability to ±0.0007 °C
  • Independent high- and low-temperature cutout circuit
  • Minimal long-term drift

 

Specs

 

Specifications

7015 7009 7108
Range 0 °C to 50 °C 0 °C to 50 °C 20 °C to 30 °C
Stability at 25 °C ±0.0007 °C (water)
±0.001 °C (mineral oil 5011)
±0.002 °C (water)
±0.004 °C (mineral oil 5011)
Uniformity ±0.003 °C at 25 °C (water)
±0.005 °C at 25 °C (mineral oil 5011)
±0.005 °C (water)
±0.008 °C (mineral oil 5011)
Temperature Setting Digital display with push-button data entry
Set-Point Resolution 0.002 °C; high-resolution mode, 0.00003 °C
Display Resolution 0.01 °C
Digital Setting Accuracy ±1 °C ±0.5 °C
Digital Setting Repeatability ±0.01 °C
Heaters 500 and 1000 Watts Peltier heating/cooling
Cooling Capacity 100 to 200 Watts 100 W in ambient 23 °C
Access Opening 699 x 279 mm (27.5 x 11 in) 699 x 559 mm (27.5 x 22 in) 356 x 356 mm (14 x 14 in)
Bath Chamber Dimensions (HxWxD) (unobstructed space) 699 x 279 x 330 mm
(27.5 x 11 x 13 in)
669 x 559 x 330 mm
(27.5 x 22 x 13 in)
356 x 203 x 355 mm
(14 x 8 x 14 in)
Depth 330 mm (13 in) 203 mm (8 in)
Wetted Parts 304 stainless steel Tank: 304 stainless steel
Resistor rack: hard-anodized, perforated aluminum
Safety Cutout Factory-set high temperature n/a
Power 115 VAC (±10 %), 60 Hz, 15 A or 230 VAC, 50 or 60 Hz, 8 A, specify 230 VAC (±10 %), 50 or 60 Hz, 12 A (specify frequency) 115 VAC (±10 %), 50/60 Hz, 3 A or 230 VAC (±10 %), 50/60 Hz, 1.6 A, specify
Volume 95 liters (25 gallons) 167 liters (44 gallons) 51 liters (13.2 gallons)
Weight 141 kg (310 lb.) 150 kg (330 lb.) 35 kg (75 lb.)
Size (HxWxD) 1219 x 1118 x 559 mm
(48 x 44 x 22 in)
1092 x 1130 x 864 mm
(43 x 44.5 x 34 in)
489 x 413 x 559 mm
(19.25 x 22 x 25 in)
Automation Package Interface-it software and RS-232 computer interface are included for setting the bath temperature via an external computer. For IEEE-488, add the 2001-IEEE.
Although the 7015 and 7009 baths are capable of reaching higher temperatures, they are designed for use with standard resistors. Therefore, the soft cutout of the instrument has been set at the factory to 50°C to protect standard resistors placed in the bath.

 